Lights and shadows of implantable cardioverter defibrillators implantations in hypertrophic, dilated and arrhythmogenic right ventricular cardiomyopathies.

In the past decades, implantable cardioverter defibrillators (ICDs) have definitively proved to be superior to antiarrhythmic drugs for the prevention of arrhythmic sudden cardiac death (SCD). Several studies have proved this hypothesis both for primary or secondary (i.e. after a sustained ventricular tachycardia with hemodynamic compromise or ventricular fibrillation) prevention of SCD. However, some complications often occur with ICD implantation, which may be particularly serious for young individuals.

Indications for ICD implantation in young people are mostly ion channel diseases and cardiomyopathies. The latter are structural abnormalities of the myocardium, related to genetic abnormalities, predisposing to malignant ventricular arrhythmias and requiring ICD for the prevention of SCD in some cases.

Hypertrophic cardiomyopathy

Hypertrophic cardiomyopathy (HCM) is one of the most common cardiomyopathies and occurs in the 0.2% of the general population1. It is characterized by localized asymmetrical hypertrophy of the heart and disarray of myocardial cells and filaments. This leads to impaired transmission of electrophysiologic impulses, with an unstable electrical substrate, potentially causing lethal ventricular tachyarrhythmias and SCD.

Hypertrophic cardiomyopathy

Short axis view of a cardiac magnetic resonance imaging of a patient affected by hypertrophic cardiomyopathy involving the interventricular septum and the anterior wall of the left ventricle.

ICD implantation in HCM: lights

A minority of patients with HCM are judged to be at increased risk for SCD, whose rate is estimated to be about 1% per year 2. However, malignant ventricular arrhythmias remains the most frequent cause of death in this population.

Beta-blockers have failed to demonstrate significant protection from SCD in HCM patients. Type I and III anti-arrhythmic drugs, including amiodarone, have been abandoned because of inefficacy and pro-arrhythmic side effects. Experiences with endocardial and epicardial mapping and ablation in HCM are poor and limited to highly selected patients.

ICDs, conversely, have been proved effective in terminating life-threatening ventricular tachyarrhythmias in HCM and are currently considered the only available tools altering the natural course of the disease and prolonging life.

Appropriate intervention rates of ICD in HCM have been esteemed to be about 11% and 4% for secondary and primary prevention respectively3.

Current recommendation for ICD in HCM

HCM has heterogeneous clinical manifestations in different individuals and a not well predictable clinical course. Targeting HCM-patients for prophylactic ICD implantation can therefore be challenging. However, some “risk factors” have been individuated, that commonly guide the decision for ICD implantation. Current international guidelines recommend ICD in all cases of secondary prevention or when familiar history of SCD, marked left ventricular (LV) hypertrophy or recent unexplained syncopal episodes are present. The role of ICD is uncertain for non-sustained VT or an abnormal blood pressure response with exercise.

ICD implantation in HCM: shadows

Previous studies have reported up to 5.1%/year rate of ICD-related complications in HCM4.

The most frequently reported is inappropriate shock delivery, mostly due to erroneous detection and treatment of atrial fibrillation by the ICD. A retrospective study on 334 consecutive HCM patients with an ICD, showed a rate of appropriate vs. inappropriate shock delivery of 2.3 and 4.6%/year5. Moreover, a recent meta-analysis involving 2190 ICD-recipients affected by HCM, showed a rate of appropriate vs. inappropriate ICD interventions of 3.3 vs. 4.8%/year respectively4.

Another extremely rare complication with ICDs in HCM, is represented by loss of capture due to high pacing thresholds (i.e. energy required for effective right ventricular pacing). This is related to ventricular hypertrophy and may be prevented by accurate individuation of the optimal pacing threshold during ICD programming.

Lengthened mitral valve leaflets, and left ventricular outflow tract obstruction due to LV hypertrophy, also confers a relatively higher risk of infection and endocarditis to HCM patients, which must be considered when an ICD implantation is programmed. Moreover, since ICD implantation in HCM is commonly performed at a relatively young age, patients are supposed to necessitate of a number of interventions for pulse generator substitution over their entire life. For this reason, risk of infection and endocarditis (related to ICD substitutions) is amplified. Eventual needing of ICD and intra-cardiac leads extraction over time must be considered. This is a high-risk procedure, possibly complicated by cardiac tamponade, shock, anemia, arrhythmias or even death.

Finally, ICD implantation can be accompanied by depression, anxiety, reduced quality of life, particularly in young people.

Idiopathic dilated cardiomyopathy

Idiopathic dilated cardiomyopathy (IDCM) is a myocardial disease characterized by LV dilation and systolic dysfunction, commonly resulting in heart failure (HF) and for whom an etiological basis cannot be identified. IDCM is relatively rare (36.5% new cases/year/100.000 persons) but accounts for nearly 10.000 deaths/year in the United States, both due to HF and arrhythmic SCD6.

Idiopathic dilated cardiomyopathy.

Long axis view of a cardiac magnetic resonance imaging of a patient affected by dilated  cardiomyopathy

ICD implantation in IDCM: lights

ICD are effective for the prevention of SCD in IDCM and can favorably alter the natural course of the disease. Previous studies have showed a rate of appropriate interventions of 5 to 7.5%/year in IDCM-patients with an ICD implanted for primary prevention 7-8. A large meta-analysis has showed a reduction of mortality with ICD of about 3.5%/year compared to the best medical therapy in IDCM 9.

Recommendation for ICD in IDCM

The most important risk factor for ventricular arrhythmias in primary prevention of SCD in IDCM, is represented by severe contractile dysfunction, quantified by the measurement of the ejection fraction (EF) at echocardiography. Functional status, assessed by the New York Heart Association (NYHA) class, also plays a role.

Current international guidelines recommend ICD always for secondary prevention. In case of primary prevention, ICD must be implanted in patients with LVEF ≤35%  and NYHA class II and III10. However, in a real-life setting, the decision making for patients with IDCM is more complex, because issues such as additional cardiac resynchronization therapy (CRT), co-morbidities, the potential to improve LVEF over time, and eventual genetic etiology also should be considered. With regard to the last issue, carriers of mutations in gene encoding for laminin, are tough to be at increased risk of complete atrio-ventricular block, malignant ventricular arrhythmias and SCD, and may therefore require ICD implantation independently from the EF%.

ICD implantation in IDCM: shadows and warnings

Some shadows obscure the scenario of SCD prevention with ICDs in IDCM.

An important issue regards the high number of patients to be treated in order to save one life because of a currently adopted risk stratification process that appears to lack specificity.

Another topic is the correct timing for ICD implantation for primary prevention. A post hoc analysis of the DEFINITE trials 11 showed that only patients who had received their ICD not later than 3 months after the diagnosis of IDCM would have benefit from implant. However, a significant number of IDCM-individuals show marked improvement of the EF over time, up to values higher than those for which ICD implantation is currently recommended. The goal therefore, appears to be an early and correct individuation of those subjects who will have a negative clinical course, and will therefore require an early ICD implantation for primary prevention.

Similarly to HCM, ICD-related complications have been also reported by various studies, among whom:

  1. inappropriate shocks, with consequent reduction of quality of life;
  2. infections;
  3. risks related to eventual lead extractions;
  4. depression and anxiety.

Arrhythmogenic right ventricular cardiomyopathy

Arrhythmogenic right ventricular cardiomyopathy (ARVC) is a genetic cardiomyopathy characterized by progressive fibro-fatty replacement of the right ventricular myocardium, sometimes extending also to the left ventricle. The clinical presentation is usually related to ventricular tachycardia (VT) with a left bundle branch block pattern or ventricular fibrillation (VF) leading to SCD. ARVC is a progressive disease ultimately leading to HF.

03-PanelA

  Panel A

04-PanelB

Panel B

Short (Panel A) and long  axis views (panel B) of a cardiac magnetic resonance imaging of a patient affected by arrhythmogenic right ventricular cardiomyopathy, with evident fibro-fatty replacement of the right ventricle wall, the apex and partial involvement of the left ventricle wall.

ICD implantation in ARVC: lights

The therapeutic options for ventricular arrhythmias in ARVC include catheter ablation and antiarrhythmic drugs such as beta blockers, sotalol and amiodarone. However, these strategies have proved to improve symptoms but not to increase survival. Nowadays, ICD remains the most effective safe-guard against SCD in ARVC.

In a large multicenter international study enrolling ARVC patients with an ICD implanted for primary prevention, appropriate device interventions were observed in one fourth of patients after 5 years. The annual rate of potentially “life-saving” shocks against VF was 3.3% and the estimated benefit of ICD implantation was of 23% after 2 years 12. Another recent prospective study enrolling a cohort of ARVC patients with an ICD implanted for primary prevention, reported appropriate device interventions in nearly one-half of individuals over a period of 4.7 years 13.

Recommendation for ICD in ARVC

As for IDCM and HCM, ICD implantation is recommended in ARVC for secondary prevention (aborted SCD, VF or hemodynamically unstable sustained VT). ICD is also recommended for primary prevention in patients with LVEF≤35%, severe right ventricular dilation and/or dysfunction, a syncopal episode suggestive for VT or VF or an affected family member with SCD. Other risk factors include: non-sustained VT, early onset of the disease and competitive sport activity.

ICD implantation in ARVC: shadows.

Several studies have proved a relatively high incidence of device-related complications in ARVC patients with an implanted ICD. In a study of 132 patients, five individuals required an additional lead because of pacing failure and one patient died from endocarditis secondary to device infection 14. Progression of fibro-fatty replacement of the myocardial tissue has been associated with high pacing thresholds and impedances and eventual loss of capture. Up to 37% lead-related complication in 7 years have been described in ARVC patients with an implanted ICD 15.

Considerations

The decision regarding ICD implantation is highly significant for any individual at risk of SCD. The clinical decision-making process itself is complex and imply consideration of a number of different aspects. Despite ICD is the only life-prolonging therapy in cardiomyopathies, eventually associated complications should be considered.

Modern technology, such as that utilized by NayaMed, has currently developed sophisticated algorithms allowing an accurate discrimination of supraventricular from ventricular arrhythmic episodes and properly guiding ICD therapies. A growing knowledge of those algorithms by the cardiologists would hopefully reduce the rate of inappropriately delivered shocks by ICDs.

Reliable RV lead alerts and RV lead integrity algorithms, also a constant automatic measurement of leads’ impedances and P, R Waves amplitude have also been implemented in the ICDs of NayaMed, preventing the problems caused by a lead failure. However, despite such algorithms, a complete abolition of ICD-related complications in cardiomyopathies is unreliable. Therefore, an accurate risk stratification appears necessary, as well as a careful weigh of the relative risks and benefits of ICD implantation in each individual.

Dr. Annamaria Martino
Dr. Leonardo Calo
Policlinico Casilino
Roma, Italy

References

1)      Gersh J, Maron BJ, Bonow RO et al. 2011 ACCF/AHA Guideline for the Diagnosis and Treatment of Hypertrophic Cardiomyopathy: A Report of the American College of Cardiology Foundation/ American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol 2011; 58:e212-260

2)      Maron BJ. Hypertrophic cardiomyopathy: a systematic review. JAMA 2002; 287:1308-1320

3)      Maron BJ, Spirito B, Shen WK et al. Implantable cardioverter defibrillators  and prevention of sudden cardiac death in hypertrophyc cardiomyopathy. JAMA 2007; 298: 405-412

4)      O’Mahony C, Lambiase PD, Quarta G et al. The long-term survival and the risk and benefits of implantable cardioverter defibrillators in patients with hypertrophic cardiomyopathy. Heart 2012;98:116-125

5)      Schinkel AF, Vriesendorp PA, Sijbrands EJ et al. Outcome and complications after implantable cardioverter defibrillator therapy in hypertrophic cardiomyopathy: systematic review and meta-analysis. Circ Heart Fail 2012;5:552-559

6)      Codd MB, Sugrue DD, Gersh BJ et al.  Epidemiology of idiopathic dilated and hypertrophic cardiomyopathy. A population based study in Olmsted County, Minnesota, 1975–1984. Circulation 1989;80:564–72

7)      Bardy GH, Lee KL, Mark DB, Poole JE et al. Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352:225–237.

8)      Kadish A, Dyer A, Daubert JP et al; Defibrillators in non-ischemic Cardiomyopathy Treatment Evaluation (DEFINITE) Investigators. Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy. N Engl J Med 2004;350:2151–2158.

9)      Desai AS, Fang JC, Maisel WH, Baughman KL. Implantable defibrillators for the prevention of mortality in patients with nonischemic cardiomyopathy: a meta-analysis of randomized controlled trial. JAMA 2004;292: 2874–287

10)  Zipes DP; Camm AJ, Borggrefe M et al. ACC/AHA/ESC 2006. Guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Europace 2006; 8:746-837

11)  Kadish A, Schaechter A, Subacius H et al. Patients with recently  diagnosed nonischemic cardiomyopathy benefit from implantable cardioverter defibrillators. J Am Coll Cardiol 2006;47:2477–2482

12)  Corrado D et al. Prophylactic Implantable Defibrillator in Patients With Arrhythmogenic Right Ventricular Cardiomyopathy/Dysplasia and No Prior Ventricular Fibrillation or Sustained Ventricular Tachycardia. Circulation. 2010;122:1144-1152.

13)  Bhonsale A, James CA, Tichnell C  et al. Incidence and predictors of implantable cardioverter-defibrillator therapy in patients with arrhythmogenic right ventricular dysplasia/ cardiomyopathy undergoing implantable cardioverter-defibrillator implantation for primary prevention. J Am Coll Cardiol, 2011;58:1485-96

14)  Corrado D, Leoni L, Link MS et al. Implantable cardioverter-defibrillator therapy for prevention of sudden death in patients with arrhythmogenic right ventricular cardiomyopathy/dysplasia. Circulation 2003;108:3084–91.

15)  Wichter T, Paul M,Wollmann C et al. Implantable cardioverter/defibrillator therapy in arrhythmogenic right ventricular cardiomyopathy. Single-center experience of long-term follow-up and complications in 60 patients. Circulation 2004;109:1503–8.


Audible alerts or How to make the ICD more efficient and the patient safer

An ICD is saving lives. It gives peace of mind to our patients thinking that at any moment, if their heart stops “someone” will be there to “bring it back”. Taking in consideration that most people who experience a sudden cardiac arrest do not survive the first event, the ICD is one of the great medical inventions of the 20th century.

In order for the ICD to treat the ventricular arrhythmia it needs to accurately detect them. Undersensing and Oversensing are the two significant problems affecting the detection in an ICD.

There are also 2 categories of situations in which the patient should be notified: if his clinical condition is getting worse and if the system’s functionality is in danger. In big majority of cases the ICDs are reliable devices but, sometimes a part of the system can break and in those moments it is extremely important to warn the patient immediately in order for him to contact his physician as soon as possible. This is particularly true when the integrity of the defibrillation lead is threatened.

Soon after the implantable defibrillator became the 1st line of defence against the SCD, physicians realised the inappropriate therapies are the biggest complication.

Getting an ICD is an important moment in a patient’s life and for primary prevention it is something difficult for the patient to understand “Why do I need a device when I never had a cardiac arrhythmia before?” Now, imagine this patient after getting his primary prevention ICD starting a home activity where he is using his arm a little more than before, like playing tennis with his son.  Imagine that after a few days he starts receiving inappropriate shocks. He’s been told the ICD is there to save his life but in exchange he’s passing in pain because of it. This is just an unfortunate event but it is the reason why in NayaMed we have implemented a number of clinical and device alerts specifically designed to warn the patients about potential problems.

The ICD leads are complex medical devices and all the manufacturers are seriously testing them before the market release but with all this, a problem can appear. In those situations the NayaMed ICD will start ringing an alarm at constant intervals and it is among the only devices in the market that will ring daily, until a physician is checking the device.

As the right ventricular lead problems represent the major complication in the ICD, NayaMed has 3 separate analysis and alerts aiming to detect a possible RV lead problem.

The constant impedance check is the classical alert that will ring if the impedance value goes out of certain ranges, but NayaMed ICD can do much more…

point

The Lead Integrity Alert is probably one of the most ingenious and useful alerts on the market. It is monitoring the important variations in the lead impedance, the short not-physiological ventricular intervals (close to the ventricular blanking) and highly rapid non-sustained ventricular events (that could be due to a lead rupture and not to real heart contractions).  It integrates the information from those 3 parameters in the following way:

  • An RV pacing lead impedance measurement is less than 50% or greater than 175% of the baseline impedance. The baseline measurement is the median of the previous 13 daily measurements.
  • The ventricular Sensing Integrity Counter is incremented by at least 30 within a period of 3 consecutive days or less.
  • The device senses 2 High Rate-NS episodes with a 4-beat average R-R interval of less than 220 ms.

If at least 2 of the following criteria have been met within the past 60 days, it identifies a potential lead fracture.

graph_article

When the criteria for the RV Lead Integrity Alert are met, the device immediately sounds an alert tone. The tone sounds again every 4 hours beginning at the next scheduled 4-hour time interval. The tone also sounds at the programmed Alert Time and when a magnet is placed over the device.

What is important is in the NayaMed ICD, the tone continues to sound until the device is interrogated by a programmer.

Using this alert, in case of a lead fracture, the patient will receive a 3 days warning* before the inappropriate shocks start. As mentioned in the graph, it is not a simple alert but it also takes some proactive actions to minimize the consequences of a broken lead and help to diagnose it.

point

One of the newest alerts on the device market is the RV lead noise which is actually differentiating between noise and a real VT/VF event. When VT/VF detection is withheld due to lead noise and the RV Lead Noise Alert is on, the device sounds a patient alert tone. This tone then sounds every 4 hours beginning at the next scheduled 4-hour time interval. The alert tone also sounds at the programmed Alert Time and when a magnet is placed over the device.

noise_article
The Important aspect… the alert tone continues to sound until the device is interrogated by a programmer. As the previous one, it is not only an alert but also a discriminator preventing the inappropriate shocks due to noise.

VF Detection or VF Therapy OFF, low battery voltage and the excessive charge time are also 3 alerts warning the patient that he/she should consult his physician as soon as possible.

All the device alerts are nominally switch ON so basically we can say that if the system has a hardware problem… the patient will be notified.

The physician also has the possibility to program the alert time in respect with patient preferences and activity.

point

In addition to those important device alerts there are 2 clinical alerts helping the physicians to better manage the patients; in case the number of shocks delivered in a VT/VF episode is greater than or equal to the programmed Number of Shocks Threshold and if a specific VF, VT, or FVT episode was redetected after all programmed therapies for that type of episode were delivered.

All the alerts can be easily programmed from the Alert button on the Parameter screen.

programmer_article

Analysing all those aspects we’re happy to say that the NayaMed ICD has probably one of the biggest range of audible device alerts that will constantly monitor the system and ring when necessary. They work not only with the NayaMed leads but with any compatible lead on the market making the ICD a better lifesaver.

*Swerdlow CD, Gunderson BD, Ousdigian KT, Abeyratne A, Sachanandani H, Ellenbogen KA. Downloadable Software Algorithm Reduces Inappropriate Shocks Caused by Implantable Cardioverter-Defibrillator Lead Fractures. A Prospective Study. Circulation. Published online September 27, 2010. 

Alexandru Trif


Thank you for visiting us at the ESC Congress 2012 in Munich


NayaMed products in a few words

As a new appearance on the CRDM Market, customers often ask: “Who are you NayaMed?”, “What are your products and why should we chose you for our patients?”

Well, mostly because we offer a simple online platform with a permanent access to your products, giving full transparency and allowing permanent access to ordering, digital inventory, stock management access and technical support.

Also, because as a clinician, you get the product you need when you need. A product that has been designed and engineered to perfectly match the NayaMed philosophy – Smart and Simple.

We wanted the clinicians to be comfortable when using our devices, so we asked our Technical Advisors to present the NayaMed products in just a few minutes.

Mary Soranno and Marina Mancusi, our technical advisors for the UK and the Italian market did a wonderful job and we are inviting you to see a very brief, but complete presentation of our products in English or Italian.

English Version:

Italian Version:


NayaMed devices made simple. Watch it now!

Meet Mary Soranno, our technical advisor for the UK market presenting all the features of the NayaMed pacemakers & defibrillators. (Better than a long blog post or brochure!).


CARDIOSTIM 2012

Cardiostim_2012

Cardiostim 2012, the 18th world congress in cardiac electrophysiology and cardiac techniques has been held in Nice last week and this time more than 5.000 delegates coming from over 100 countries have crowded the congress halls in the Acropolis.

Industry representatives and clinicians had the possibility to meet here in order to discuss on clinical outcomes, new techniques and future technologies.

One of the major innovations presented at the congress had been the 3D mapping and navigation systems dedicated to diagnostics and interventional rhythmology.

These revolutionary tools have been designed to help the implantation of stimulation catheters in CHF patients in order to obtain the best results from the cardiac resynchronization, to simplify the procedures and to reduce the use of fluoroscopy.

In terms of products, Medtronic presented CardioGuide a navigation system for cardiac resynchronization therapy implants. The system is able to generate 3D images of coronary vessels allowing a better road map for finding the best location for the LV lead placement.

St Jude Medical presented MediGuide a mini 3D GPS mapping system able to visualize in real-time the location of a MediGuide enabled device.

A strong focus was given also to the new MRI conditional devices.

Biotronik was present with a big booth entirely dedicated to proMRI devices and in particular to the new Lumax 740 proMRI the first ICD MRI conditional allowing for the first time tachyarrhythmia patients fitted with a device to undergo MR scans.

St. Jude Medical was promoting the Accent MRI family of pacemakers highlighting the availability of a dedicated MRI activator to simplify the device pre-MRI procedures.

Boston scientific showed a panel with a picture of the new pacemaker Ingenio MRI even if this new model MRI compatible seems not yet approved.

A good interest was around new CRT optimization systems with the purpose to improve the response of heart failure patients indicated for CRT.

Sorin presented the new SonR System based on a new hemodynamic sensor mounted in the tip of a dedicated atrial lead (SonRtip) that in combination with the new CRTD device (Paradym RF SonR) is able to provide an automatic weekly optimization.

In the same direction St Jude Medical was promoting the already introduced Quadripolar CRT system applied to the Quadra family of ICDs. This technology enables the physician to find the best pacing site choosing between 10 different LV pacing vectors.

Cameron Health (recently acquired by Boston Scientific) was present at the congress with a booth dedicated to the new S-ICD System, the world’s only ICD that provides reliable defibrillation therapy without transvenous leads. Unlike traditional ICDs that require leads placed in the hearth, this system is the first implantable completely subcutaneously leaving vasculature untouched and eliminating completely complications associated with transvenous leads. In fact both the device and the electrode are implanted just under the skin and outside the ribcage.

In a similar direction Medtronic seems to be working to realize a leadless pacemaker with reduced dimensions that will be implanted directly inside the hearth.

Between other topics, St Jude Medical had a part of the booth dedicated to new Ellipse ICD that with 31cc  of volume and 66g of weight seems to be the smallest ICD available on the market (even if not very far from the latest models of Medtronic and Boston Scientific).

Boston Scientific and Sorin were also proposing diagnostics solutions or clinical studies dedicated to the sleep apnea syndrome (SAS), the most widespread sleep breathing disorder commonly associated with increased cardiovascular morbidity and mortality.

The Cardiostim Innovation Award, selected by the congress attendees, honored the S-ICD of Cameron Health, the SonR optimization system of Sorin Group and the CardioGuide Implant System of Medtronic.

NayaMed, the new born cardiac device company with a strong focus on customer optimization and processes simplification, was not present at the congress.

Was it a way to differentiate, again, from all the other companies or just a matter of time to see NayaMed booth at a congress?”

In conclusion, in the beautiful contest of the city of Nice, a very crowded congress with interesting new technologies already available or next to come.

Be ready for the next Cardiostim appointment in June 2014… with NayaMed ?

 

Here is a quick look of this event :

Sergio Datteri, Independent Medical Device Consultant


First NayaMed ICD implant performed only with remote support

“I know that the patient is ready for the implant, but I’m late because of the snow, I’m so sorry!”;

“Hello Dr.Y, I’ll do my best to be here as soon as possible for the first scheduled procedure but I’m afraid that I won’t be able to arrive before one hour”…

How many times the “traditional” technical advisors have already had to face such embarrassing situations? And how many times do the physicians have to change the EP lab program because they need a technical support?  Really often I can say…but nor today and in the future with NayaMed!

Today at one of our customer in Italy, a NayaMed implant was performed ONLY with a remote support, without anyrepresentative in the field!!!

At the beginning of the procedure, I was in contact with the nurse online. She had performed all the initial tests on the ICD autonomously, just looking at NayaMed ICD Job aids and she asked me only few suggestions for the procedure.

At the end of the implant, the nurse called me back and we discussed together with the physician about how to optimize the parameters for our patient. It was a great and efficient work: I was available for them despite the snow at their time, even after some last-minute changes in their activity in the EP lab. Everything was done exactly in the NayaMed spirit: leading our customers to become autonomous in implanting conventional devices, being aware of having an Online Technical Support that is always available and ready for them, anytime, anywhere.

This is really a new frontier of smart and simple healthcare isn’t it?

Marina Mancusi
NayaMed Technical Advisor